Why Sub 1 nm Sizing is Important?

The requirements for dynamic light scattering (DLS) systems are usually dependent on the sample. However, numerous DLS systems require the lower size limit to be either 1 nm or less (0.3 nm). This article provides verification that the Nicomp® DLS system is proficient in taking repeated measurements of particle sizes less than 1 nm and offers specifics on the measurements’ completion.

Introduction

Dynamic light scattering is the ideal method of measurement for particle size analysis of less than 1 µm. The standard dynamic range for this technique is around 1 nm–1 µm; however, the performance at both ends depends on the sample and system. The upper size limit of DLS is subject to the volume and density of the dispersed phase.

Emulsion samples may be achievable at ~10 µm because the density of the dispersed and continuous phase has a level of similarity. However, for a high density solid particle like gold (19.29 g/cc), the upper limit will be significantly less. After the particle motion is dominated by sedimentation, instead of Brownian Motion, the result can no longer be relied upon.

The upper size limit of DLS is also impeded by number fluctuations as the small number of large particles enter and vacate the inspection area. On the small size, the specification relies on the construction and optical quality of the instrument, in addition to the scattering properties and concentration of the sample1. Measurements approaching 1 nm are difficult and need cautious sample preparation.2 This article shows measurements taken at or less than 1 nm using sucrose and the sample. Note: Optimum Nicomp performance close to 1 nm usually needs the 35 mW laser and APD detector.

Sucrose molecule.

Figure 1. Sucrose molecule. Image Credit: Entegris

Materials

Sucrose (Figure 1) was bought from Fisher Chemical, certified ACS crystalline, S5-500, lot 173322, chemical formula C12H22O11, molecular weight = 342.2965 g/mol. Sucrose solution was readied at 10 volume percent utilizing filtered DI water. All measurements were produced employing the Nicomp Z3000 DLS particle size analyzer (Figure 2) configured with a 35 mW laser diode and APD detector.

Nicomp Z3000 DLS system.

Figure 2. Nicomp Z3000 DLS system. Image Credit: Entegris

Instrument Settings

In Figure 3, the standard measurement settings used in this study can be seen.

. .
Refractive index 1.333
Viscosity 1.172 cP
Auto sensitivity Checked
Auto baseline adjustment Checked
Channel width 5 µsec
Temperature 23 °C
Intensity set point 300 kHz
First channel used 2
Scattering angle 90°
Cell type Round
Run time 7 minutes

 

Figure 3. Nicomp measurement settings. Image Credit: Entegris

Note: The viscosity data used was 1.172 cP for 10% sucrose. Precise viscosity values are necessary for correct DLS results for non-dilute based measurements.

Results: Gaussian

The Gaussian intensity and volume results3 for 10% sucrose can be seen in Figure 4.

Intensity and volume Gaussian results for 10% sucrose.

Intensity and volume Gaussian results for 10% sucrose.

Figure 4. Intensity and volume Gaussian results for 10% sucrose. Image Credit: Entegris

Results: Nicomp

The Nicomp results are established on a non-negative least squares algorithm enhanced to resolve multi-modal distributions. All results in this study produced high Chi-square3 values, suggesting the occurrence of multiple modes. The Nicomp intensity result for the 10% sucrose can be seen in Figure 5.

Intensity Nicomp result for 10% sucrose.

Figure 5. Intensity Nicomp result for 10% sucrose. Image Credit: Entegris

Conclusions

The Nicomp DLS system is well-adept at measuring as low as and lower than 1 nm, strengthening the lower size specification of 0.3 nm. The bigger second peak at approximately 100 – 200 nm has been described in other published literature4 and is considered to be impurities from the raw material and handling.

References

  1. Entegris Technical Note, 0.1 mg/mL Lysozyme
  2. Entegris Technical Note, DLS Sample Preparation
  3. Entegris Technical Note, DLS Data Interpretation
  4. Weinbuch et al., nanoparticulate impurities in pharmaceutical-grade sugars and their interference with light scattering-based analysis of protein formulations, pharm res (2015) 32:2419–2427

This information has been sourced, reviewed and adapted from materials provided by Entegris.

For more information on this source, please visit Entegris.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Entegris. (2024, February 19). Why Sub 1 nm Sizing is Important?. AZoM. Retrieved on November 21, 2024 from https://www.azom.com/article.aspx?ArticleID=19869.

  • MLA

    Entegris. "Why Sub 1 nm Sizing is Important?". AZoM. 21 November 2024. <https://www.azom.com/article.aspx?ArticleID=19869>.

  • Chicago

    Entegris. "Why Sub 1 nm Sizing is Important?". AZoM. https://www.azom.com/article.aspx?ArticleID=19869. (accessed November 21, 2024).

  • Harvard

    Entegris. 2024. Why Sub 1 nm Sizing is Important?. AZoM, viewed 21 November 2024, https://www.azom.com/article.aspx?ArticleID=19869.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.