Analyzing Infant Cereals

This article shows how the Epsilon 3XLE benchtop XRF spectrometer is suitable for analyzing Na, Mg, P, Ca, Mn, Fe, Zn, Cl, and K elements in infant cereals prepared as loose powders or pressed pellets.

Instrumentation

A Panalytical Epsilon 3XLE EDXRF spectrometer was used to perform the measurements. This spectrometer is equipped with a high-resolution SDD-10 Detector, 50 kV 3 mA Rhodium anode X-ray tube, helium purge, spinner, 6 filters and a 10-position removable sample changer.

Standards

For this study, a set of 8 infant cereal powder standards comprising of Na, Mg, P, Ca, Mn, Fe, Zn, Cl and K was used. A potentiometric method for (Cl) and ICP-AES (all other elements) was previously used to determine elemental concentrations.

Sample Preparation

Pressed Pellets

A 7.5 g sample from each standard was pressed at 5 or 10 t (high fat or plant content, respectively) for 90 seconds using an automatic Herzog HTP 40 press together with a 40 mm WC pressure die to prevent Fe contamination.

Loose Powders

Next, 6 grams of each sample was transferred into a (P1) sample cup assembled with a high transmission Prolene (4 µm) supporting foil. Then, all powders were lightly compressed with a hand stamp tool.

Measurement Procedure

As shown in Table 1, four different range optimized measurement conditions were used. All analyses were conducted in a helium/air environment and the total counting time was 15 minutes. A region-of-interest (ROI) method was used to measure the intensities for Mg and Na, while all other elements were measured using completely deconvoluted spectra.

Table 1. Measurement conditions

Condition kV µA Measurement time (s) Medium Filter Elements
<F-Si> 5 3000 600 Helium None Na, Mg, P
<K-V> 12 1250 120 Air Al-thin Cl, K, Ca
<Cr-Co> 20 360 120 Air Al-thick Mn, Fe
<Ni-Mo> 50 170 60 Air Ag Zn

Calibration Results

Standards were measured in triplicate. Calibration plots for K and Na are shown in Figures 1 to 4. A good correlation was observed between the chemical concentrations and the measured intensities. The detection limits (LLD) and calibration data for both preparation methods are summarized in Table 2.

Table 2. Pressed pellet and loose powder calibration results

Compound Concentration range (mg/100 g) RMS* (mg/100 g) Correlation LLD (3σ) (mg/100 g, appl. time)
Pressed pellets Loose powder Pressed pellets Loose powder Pressed pellets Loose powder
Na 6.00 - 276.50 14.28 17.62 0.9908 0.9867 14.3 71.4
Mg 16.00 - 137.50 7.79 7.13 0.9837 0.9864 5.0 25.0
P 119.50 - 557.50 13.93 16.87 0.9955 0.9934 2.0 2.5
Cl 24.50 - 454.00 15.20 14.17 0.9958 0.9964 2.0 2.0
K 84.00 - 942.50 17.96 19.61 0.9977 0.9973 2.0 2.0
Ca 163.50 - 629.00 18.26 22.10 0.9931 0.9899 1.2 1.2
Mn 0.13 - 3.40 0.07 0.06 0.9978 0.9980 0.2 0.2
Fe 4.36 - 26.30 0.97 1.06 0.9895 0.9872 0.2 0.2
Zn 1.26 - 5.45 0.52 0.74 0.9379 0.8697 0.2 0.2

* RMS (Root Mean Square) is an indication of the quality of the regression. The more accurate calibrations have smaller RMS values.

Calibration graph for Na in pressed pellets.

Figure 1. Calibration graph for Na in pressed pellets.

Calibration graph for Na in loose powders.

Figure 2. Calibration graph for Na in loose powders.

Calibration graph for K in pressed pellets.

Figure 3. Calibration graph for K in pressed pellets.

Calibration graph for K in loose powders.

Figure 4. Calibration graph for K in loose powders.

Precision

In order to test the stability of the instrument, one pelletized sample was repeatedly measured for 20 times. Table 3 shows the average concentration, the relative RMS, and RMS (1σ) of the repeat measurements. All elements produced a relative RMS value < 1.65 %, except for Mn, owing to its low concentration, where counting statistics play a major role.

Compound Average concentration (mg/100 g) RMS (mg/ 100 g) Rel. RMS (%)
Na 194.79 2.35 1.21
Mg 63.15 0.63 1.00
P 414.62 1.24 0.30
Cl 438.20 2.31 0.53
K 671.67 1.55 0.23
Ca 641.14 1.49 0.23
Mn 0.15 0.02 16.53
Fe 6.54 0.05 0.69
Zn 3.47 0.06 1.61

 

Conclusion

The data presented reveals that the Epsilon 3XLE is suitable for rapid analysis of infant cereal samples. There were only slight differences between loose powder and pressed pellets preparations.

The repeatability results show the robustness and stability of the Epsilon 3XLE. The combination of high sensitivity, excellent detector resolution and powerful software deconvolution models contribute to the precision and accuracy of the results.

This information has been sourced, reviewed and adapted from materials provided by Malvern Panalytical.

For more information on this source, please visit Malvern Panalytical.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Malvern Panalytical. (2023, May 30). Analyzing Infant Cereals. AZoM. Retrieved on November 21, 2024 from https://www.azom.com/article.aspx?ArticleID=14034.

  • MLA

    Malvern Panalytical. "Analyzing Infant Cereals". AZoM. 21 November 2024. <https://www.azom.com/article.aspx?ArticleID=14034>.

  • Chicago

    Malvern Panalytical. "Analyzing Infant Cereals". AZoM. https://www.azom.com/article.aspx?ArticleID=14034. (accessed November 21, 2024).

  • Harvard

    Malvern Panalytical. 2023. Analyzing Infant Cereals. AZoM, viewed 21 November 2024, https://www.azom.com/article.aspx?ArticleID=14034.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.