Discover the Advantage of High-Resolution Quadrupole Mass Spectrometry

Process Insights’ MAX 50™ flange mounted quadrupole mass spectrometer and VeraSpec™ HRQ turnkey system integrates a 19 mm tri-filter quadrupole and 2.9 MHz Quadrupole Power Supply (QPS) to offer users the tools to deliver rapid, sensitive and consistent analysis. Accomplishing this level of performance using quadrupole mass spectrometry is an effective form of high-resolution analysis.

Image Credit: Shutterstock/Jens Goepfert

Product Advantages:

  • High resolution >4000 (M/ΔM–FWHM)
    • High resolution can be realized while scanning the whole mass range
  • Variable resolution
    • User selectable resolution across whole (50 amu) mass range
  • Excellent high-resolution mass stability
  • High-abundance sensitivity
  • Fast, high-resolution scans
    • Up to 5000 data points per amu in scans
  • Low detection limits
    • Approximately 10 ppm under high-resolution conditions

VeraSpec HRQ High Resolution Quadrupole Mass Spectrometer.

Figure 1. VeraSpec HRQ High Resolution Quadrupole Mass Spectrometer.

MAX 50 Flange Mounted Quadrupole Probe and Control System.

Figure 2. MAX 50 Flange Mounted Quadrupole Probe and Control System.

High Resolution

The resolution capabilities are >4000 (M/ΔM–FWHM), measured with Argon at m/z 40. This permits the separation of species such as helium and deuterium, 4.002 and 4.028 amu, respectively. Figure 3 presents an example spectrum. This spectrum was collected with the help of a scan speed of almost 1-second per scan.

3.98–4.04 amu Scan of 10% Helium, 10% Deuterium, Balance Argon

Figure 3. 3.98–4.04 amu Scan of 10% Helium, 10% Deuterium, Balance Argon

Variable Resolution—User Selected over Entire Mass Range

Typical use of a quadrupole mass spectrometer needs a nominal resolution setting at 1 amu across the mass range, or unit-mass resolution. Process Insights system hardware is calibrated and set in the factory in order to develop spectra with unit-mass resolution for all masses in range with zero (or near zero) software adjustments. Calibration points, or software adjustments, can be added anywhere in the table within the mass range. This enables the user to effortlessly adjust the settings to fit the requirements of their experiment.

Resolution Calibration Table.

Figure 4. Resolution Calibration Table.

Resolution points at m/z 4, 5, 18, 28, 32, and 40 were further added to the table and set to 0.0. When the resolution at m/z 28 is altered, the resolution of the peaks between m/z 28 and the adjacent calibration point will get affected. The relative adjustment is demonstrated in the resolution calibration table and the changed resolution of m/z 28 and m/z 29 is displayed in Figure 5.

Resolution Calibration—Adjusted Table.

Figure 5. Resolution Calibration—Adjusted Table.

Entire scans under high-resolution conditions, or entire scans with unit-mass resolution, can be run if specified points are not added to the table. The addition of many points to the table  produces multiple regions, each with varied resolution settings across the scan.

An instance of a complete, 1-50 amu, scan with different resolution is illustrated in Figure 6. The spectrum was recorded with the help of a two-second scan, high-resolution settings from m/z 3.9-4.1 amu, and nominal resolution elsewhere.

A mass spectrum of 1-50 amu showing a variable resolution. Sample consists of 10% helium, 10% deuterium, and balanced argon.

Figure 6. A mass spectrum of 1-50 amu showing a variable resolution. Sample consists of 10% helium, 10% deuterium, and balanced argon.

High Abundance Sensitivity

Abundance sensitivity is considered to be an accurate value representing the efficiency of how the mass spectrometer can measure adjacent peaks. This is vital for applications dealing with neighboring spectral signals with significantly different concentrations. The exceptional abundance sensitivity is shown in Figures 7 and 8.

UHP helium scan for analysis of 3 He Isotope (1.4 ppm Relative Abundance). The Resolution Was Set Appropriately for the Isotopic Analysis (Less than the maximum resolution for the scan).

Figure 7. UHP helium scan for analysis of 3 He Isotope (1.4 ppm Relative Abundance). The Resolution Was Set Appropriately for the Isotopic Analysis (Less than the maximum resolution for the scan).

High-resolution scan conditions are used for observing this abundance sensitivity. In an analysis of helium and deuterium at changing relative concentrations, a high abundance sensitivity eradicates any spectral interference between the two species. In Figure 8, the special overlay shows four such scans. The scans were collected using the following:

  • UHP helium (saturating the detector)
  • UHP helium when added to the system (before saturating the detector)
  • UHP helium mixed almost 50/50 with a helium/deuterium cylinder (10% helium, 10% deuterium, balance argon)
  • 10% helium, 10% deuterium, balance argon cylinder alone

High-resolution scans of varying relative concentrations of helium and deuterium collected using a one-second scan time, with 1500 Points/amu.

Figure 8. High-resolution scans of varying relative concentrations of helium and deuterium collected using a one-second scan time, with 1500 Points/amu.

Fast High Resolution Scans

Exceptional spectra are delivered without too much averaging and scan times. For experiments that need many data points in every single scan, it is able to gather up to 5000 data points/amu.

High-resolution scan of helium and deuterium with maximum data points (points/amu) collected using one-second scan with the maximum amount of points/amu. The scan range of m/z 3.99–4.04 Resulted in 250 data points.

Figure 9. High-resolution scan of helium and deuterium with maximum data points (points/amu) collected using one-second scan with the maximum amount of points/amu. The scan range of m/z 3.99–4.04 Resulted in 250 data points.

High-Resolution Mass Stability

Long-standing stability of a system is vital for extended analyses or experiments, and to prevent excessive time checking and recalibrating mass position. Stability becomes progressively more visible and important under high-resolution conditions. To monitor the stability, a cylinder of 10% helium, 10% deuterium, balance argon was connected and examined for 24 hours. Spectra at six-hour time intervals, starting at t=0, were used to evaluate the stability of the instrument. Spectra were collected using a 1 second scan with 5000 points/amu. A spectral overlay of the results found is shown in Figure 10. The experiment demonstrated practically no detectable mass spectral changes over a period of 24 hours.

Spectral overlay of high-resolution scans during 24 hour stability experiment.

Figure 10. Spectral overlay of high-resolution scans during 24 hour stability experiment.

Low Detection Limits

The resolution setting influences detection limits of a system. A sample dilution was used to experimentally determine low detection limits at high resolution. A cylinder of 10% helium, 10% deuterium, balance argon was diluted with UHP argon with the help of two mass flow controllers. Almost 50 cc/minute of dilution gas was used while the flow of sample was set to 8, 6, 4, 2, and 0 cc/minute. This data was used to produce a dilution curve and then calculate the low detection limits for both species. Tabulated detection limit results for helium and deuterium are presented in Figures 11 and 12 respectively, and the dilution curve in Figure 13.

m/z 4.002 10% He / 50 sccm Ar
Concentration 0.00% 0.42% 0.87% 1.36% 1.90%
Average 7.52 1509 3018 4551 6141
Standard Dev 1.65 18 27 30 40
RSD 22% 1.2% 0.9% 0.7% 0.7%
Sig / PPM   333250      
LDL(PPM)   14.9      

Figure 11. Low detection limit calculation for helium.

m/z 4.028 10% D2 / 50 sccm Ar
Concentration 0.00% 0.42% 0.87% 1.36% 1.90%
Average 9.02 2626 5163 7397 9079
Standard Dev 1.91 25 50 44 105
RSD 21% 1% 1.0% 0.6% 1.2%
Sig / PPM   560116      
LDL(PPM)   10.2      

Figure 12. Low detection limit calculation for deuterium.

Dilution curve from helium-deuterium detection limit experiment.

Figure 13. Dilution curve from helium-deuterium detection limit experiment.

This information has been sourced, reviewed and adapted from materials provided by Process Insights - Mass Spectrometry.

For more information on this source, please visit Process Insights – Mass Spectrometers.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Process Insights - Mass Spectrometry. (2023, January 11). Discover the Advantage of High-Resolution Quadrupole Mass Spectrometry. AZoM. Retrieved on November 21, 2024 from https://www.azom.com/article.aspx?ArticleID=16511.

  • MLA

    Process Insights - Mass Spectrometry. "Discover the Advantage of High-Resolution Quadrupole Mass Spectrometry". AZoM. 21 November 2024. <https://www.azom.com/article.aspx?ArticleID=16511>.

  • Chicago

    Process Insights - Mass Spectrometry. "Discover the Advantage of High-Resolution Quadrupole Mass Spectrometry". AZoM. https://www.azom.com/article.aspx?ArticleID=16511. (accessed November 21, 2024).

  • Harvard

    Process Insights - Mass Spectrometry. 2023. Discover the Advantage of High-Resolution Quadrupole Mass Spectrometry. AZoM, viewed 21 November 2024, https://www.azom.com/article.aspx?ArticleID=16511.

Tell Us What You Think

Do you have a review, update or anything you would like to add to this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.