The SPECTROstar Nano, equipped with BMG LABTECH’s ultra-fast UV/vis spectrometer, can measure absorbance over the entire spectrum (220 nm–1000 nm) in less than one second per well. Many common assay protocols are predefined, just a mouse click away, making it an ideal instrument to capture all the absorbance assays in a microplate or cuvette.
Features
- High-speed absorbance measurements
- Measurements in up to 1536-well plates
- Integrated cuvette port
- LVis Plate compatibility
- Incubation up to 45 °C
- Multiple shaking options
- Well scanning at up to 900 data points per well
Technical Specifications
Detection Modes
|
UV/vis absorbance spectra
|
Measurement Modes
|
Endpoint and kinetic
Spectral scanning (absorbance)
Well scanning
|
Microplate Formats
|
6 to 1536-well plates, user-definable
LVis Plate with 16 low-volume microspots (2 μL)
|
Microplate Carrier
|
Robot compatible
|
Light Sources
|
High energy xenon flash lamp
|
Detectors
|
CCD Spectrometer
|
Wavelength Selection
|
UV/vis absorbance spectrometer
Full spectra or 8 distinct wavelengths in < 1 s/well
|
Spectral Range
|
220–1000 nm
|
Sensitivity
|
Full spectrum captured in < 1 s/well
Selectable spectral resolution: 1, 2, 5, and 10 nm
OD range: 0 to 4 OD
Accuracy: < 1% at 2 OD
Precision: < 0.5% at 1 OD and < 0.8% at 2 OD
|
Read Times
|
Full spectrum from 220 to 1000 nm in less than 1 s/well
|
Shaking
|
Linear, orbital, and double-orbital with user-definable time and speed
|
Purge Gas Vent
|
System to inject an atmosphere or to pull a vacuum into the reader
|
Incubation
|
+3 °C above ambient up to 45 °C
The upper heating plate of the incubation chamber operates at 0.5 °C more than the lower plate. This prevents condensation build-up on the lid or sealer.
|
Software
|
Multi-user Reader Control and MARS Data Analysis Software included
FDA 21 CFR Part 11 compliant
|
Dimensions
|
Width: 36 cm, depth: 50 cm, height: 16 cm; weight: 10 kg
|
Optional Accessories
LVis Plate
|
Sample capacity: 16 separate microdrop wells for 2 µL samples; one standard cuvette position for up to 1 mL samples.
Quality control internal standards (optional): four NIST traceable optical density filters (approximate values of 0.1, 0.3, 0.6 and 1.0 OD); one holmium oxide filter for wavelength accuracy.
Dimensions: conforms to SBS standards for microplates.
|
THERMOstar
|
Microplate incubator and shaker
|
Upgrades
|
Please contact your local representative for upgrades including options such as detection modes, reagent injectors, etc.
|