Ultra-stable thermistor probes are incorporated into thin-walled stainless steel housings with shielded extension leads to form the Thermometrics Types S/AS/ES Temperature Standards. Every thermistor temperature standard has a sturdy design and can be immersed in liquid with the right preparation to guarantee stability over time.
Applications
- Working or secondary temperature requirements for all applications in laboratory metrology
- They are also appropriate for field application because they are not impacted by shock or vibration
- Measurements of process temperature, clinical applications, and general laboratory and hospital use
- Versions specifically designed for space and military applications
Features
- Long-term steadiness
- Sturdy structure
- Ideal for immersion in liquids
- Typically unaffected by vibration and shock
- NIST-approved calibration
Specifications
Source: Amphenol Advanced Sensors
Type |
S |
AS |
ES |
Temperature Range |
0 °C to 60 °C |
0 °C to 60 °C |
0 °C to 100 °C |
Stability |
0.005 °C/year |
0.002 °C/year |
0.005 °C/year |
Resistance vs. Temperature Characteristics |
Type |
S |
AS |
ES |
0 °C |
14250 |
11400 |
28500 |
25 °C |
5000 |
4000 |
1458 |
60 °C |
1458 |
1166 |
2915 |
100 °C |
- |
- |
925 |