Particle Analysis Using the LaserNetFines Q230 Particle Counter

Ferrous wear debris analysis on machinery lubrication oils is a key step to monitor the condition of machineries to avoid any catastrophic failure. The analysis of abnormal ferrous content in oils is a proven method.

There are magnetometer based designs like the Midas instrument and the Kittiwake Analex PQ series or time resolved dielectric instruments such as the 5200 minilab. Some instruments like the Kittiwake Analex PQ series report ferrous content in “PQ” index, a dimensionless index. The scale range of the PQ instrument is from PQ=0 up to PQ=750. The PQ reading for normal wear is < 25.

The Spectro LaserNetFines Q230 Particle Counter

The Spectro LaserNetFines Q230 Particle Counter (Figure 1) determines ferrous content for the measurement and trending of abnormal wear with the help of its advanced, patent-pending magnetometer design. With the dual magnetometer design, the Wear Classifier can measure the total ferrous content in mg/ml, ferrous particle count and size distribution for Fe > 25, 38, 50 and 100µm.

The Spectro LaserNetFines Q230 Particle Counter

Figure 1. The Spectro LaserNetFines Q230 Particle Counter

This article highlights the total ferrous measurement capability of the Ferrous Monitor instrument as well as its measurement accuracy and repeatability. The wear classifier capability or the ferrous particle count and size distribution capability of the Spectro LaserNetFines Q230 Particle Counter is not covered in this article.

Spectro Ferrous Monitor

The Ferrous Monitor in the Spectro LaserNetFines Q230 Particle Counter runs by detecting the variation in inductance of a coil caused by the introduction of ferrous material in small quantities into a sample cell within the coil (Figure 2). The two-channel instrument has a tiny coil for individual particle measurement and sizing and a larger coil to determine the total magnetic material content.

The Ferrous Monitor operates by sensing the change in inductance of a coil when a small amount of ferrous material is introduced into a sample cell inside the coil.

Figure 2. The Ferrous Monitor operates by sensing the change in inductance of a coil when a small amount of ferrous material is introduced into a sample cell inside the coil.

Experimental Procedure

This experiment evaluated the performance of a conventional magnetometer and the Q230 Ferrous Monitor against a gravimetric ferrous standard using a 50ppm solution of LNF-545, which is a Spectro Scientific gravimetric ferrous validation standard consisting of 76.2% ferrous in a 75 cSt matrix oil.

From the ferrous validation standard, six samples were taken and analyzed on both the LNF Q230 and the conventional magnetometer. After each measurement, the magnetometer was zeroed through measurement of a non ferrous sample before the next validation sample measurement. The LNF was run normally and flushed after each measurement using Electron 22.

Experimental Results

The experimental results for the six samples as well as the 50ppm reference line are presented in Figure 3.

Experimental results

Figure 3. Experimental results

The following table summarizes the results, accuracy and precision of the LNF Q230 and the conventional magnetometer:

  Q230 MIDAS
Sample 1 46.38 34
Sample 2 41.62 38
Sample 3 50.02 40
Sample 4 47.22 34
Sample 5 51.55 39
Sample 6 52.77 30
Std Dev 4.07 3.82
Avg 48.26 35.83
RSD% 8.44 10.65
Accuracy (average) 3% 28%

Conclusion

The Q230 provides more accurate and repeatable ferrous measurement when compared to the conventional magnetometer, proving its suitability for abnormal wear measurement and trending to in order to monitor the condition of machineries.

The accuracy of the Ferrous Monitor in the Q230 is ± 3%, a much better value than ± 28% accuracy of the conventional magnetometer. In addition, the Q230 shows superior repeatability with 21% less RSD% (Relative Standard Deviation) when compared to the conventional magnetometer.

This information has been sourced, reviewed and adapted from materials provided by AMETEK Spectro Scientific.

For more information on this source, please visit AMETEK Spectro Scientific.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    AMETEK Spectro Scientific. (2024, November 23). Particle Analysis Using the LaserNetFines Q230 Particle Counter. AZoM. Retrieved on December 03, 2024 from https://www.azom.com/article.aspx?ArticleID=11510.

  • MLA

    AMETEK Spectro Scientific. "Particle Analysis Using the LaserNetFines Q230 Particle Counter". AZoM. 03 December 2024. <https://www.azom.com/article.aspx?ArticleID=11510>.

  • Chicago

    AMETEK Spectro Scientific. "Particle Analysis Using the LaserNetFines Q230 Particle Counter". AZoM. https://www.azom.com/article.aspx?ArticleID=11510. (accessed December 03, 2024).

  • Harvard

    AMETEK Spectro Scientific. 2024. Particle Analysis Using the LaserNetFines Q230 Particle Counter. AZoM, viewed 03 December 2024, https://www.azom.com/article.aspx?ArticleID=11510.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.