As per standard test method ISO 13032, the analysis of sulfur in ultra-low sulfur diesel is demonstrated. Regulations worldwide have restricted the sulfur amount in a number of fuels oils with specific attention to diesel fuel. For several years road diesel has been restricted to a maximum sulfur concentration between 10-15 ppm, based on the global region.
Now, these restrictions are expanded to all diesel fuel including use in off-road diesel engines and large engines. Ultra-low sulfur highway diesel fuel (15 ppm sulfur maximum) has been used in all model year 2007 and later highway diesel vehicles and engines. It is recommended to be used in all diesel vehicles and engines. Based on the standard test method ISO 13032, the analysis of sulfur in ultra-low sulfur diesel is used.
Instrumentation
Model: |
Rigaku NEX QC+ |
X-ray tube: |
4 W Ag-anode SDD |
Sample type: |
Diesel fuel |
Film: |
Prolene |
Analysis Time: |
300 sec |
Environment: |
Helium |
Sample Ring: |
Single position 32mm |
Figure 1.
International Standard Test Method
The Rigaku NEX QC+ is compliant with ISO 13032 for the measurement of ultra-low sulfur between 8 - 50 mg/kg in diesel fuels and gasoline as described in the ISO 13032 method Scope.
Sample Preparation
For sample measurement, shake the sample bottle gently, allow bubbles to settle, and fill a 32mm XRF sample cup with 5g of the sample.
Calibration
Empirical calibration was done using a suite of 6 certified diesel calibration standards as shown in Table 1
Table 1.
Element: S
Units: ppm |
RMS Deviation: 0.76
Correlation: 0.99170 |
Sample I.D. |
Standard Value |
Calculated Value |
STD 1 |
5.0 |
5.2 |
STD 2 |
10.0 |
10.1 |
STD 3 |
15.0 |
14.0 |
STD 4 |
20.0 |
20.9 |
STD 5 |
25.0 |
24.9 |
STD 6 |
50.0 |
49.4 |
Figure 2.
Precision
Instrument repeatability (precision) is determined by ten repeat analyses of a sample in static position using a 300 sec analysis time.
Table 2.
Element: S Units: ppm |
Sample |
Standard Value |
Average Value |
Std Dev |
% Relative |
STD 2 |
10.0 |
10.2 |
0.48 |
4.8 |
STD 6 |
50.0 |
50.9 |
0.50 |
1.0 |
Detection Limit
Ten repeat analyses of a blank diesel sample without sulfur are taken with the sample being static and the standard deviation is determined. The lower limit of detection (LLD) is then defined as three times the standard deviation. The following typical LLDs are reported here using an analysis time of 300 s.
Table 3.
Element |
LLD |
Analysis Time |
S |
1.3 ppm |
300 sec |
S |
1.9 ppm |
600 sec |
Note: 1 ppm = 1 mg/kg
Conclusion
The results shown here indicate that the Rigaku NEX QC+ EDXRF analyzer offers excellent performance for the measurement of ULSD as per standard test method ISO 13032. The versatility of the NEX QC+ also makes it a suitable tool for measuring many other elements and oil matrices as well, such as lube oils, wear metals in oils, used oil and waste oil, and other elemental analysis needs throughout the petroleum industry.
This information has been sourced, reviewed and adapted from materials provided by Rigaku.
For more information on this source, please visit Rigaku.