The FOX 200 Heat Flow Meter from TA Instruments is an easy-to-use, accurate instrument for measuring thermal conductivity in compliance with ISO 8301 and ASTM C518. It can be run as a standalone unit or with PC-controlled configurations.
The FOX Heat Flow Meter provides fast, accurate results and features proprietary thin film heat flux transducers, responsive temperature control and digital thickness measurements. The FOX 200 is a versatile instrument and is particularly well-matched for measuring slices of cellular plastics.
- Easy-to-use, rapid results
- Conforms to ISO 8301 and ASTM C518
- PC-based or standalone operation
- Solid state cooling/heating
- Two Thin-Film Heat Flux Transducers
- Thickness measurement to 0.025 mm
- WinTherm software extends flexibility
- Suitable for R&D and QC testing
- Available Autosampler for unattended high-throughput productivity
Specifications
. |
. |
Maximum Sample Thickness |
51 mm (2") |
Sample Size |
204 x 204 mm (8 sq. in.) |
Temperature Range |
-20 °C to 75 °C (Fox 200)
0 °C to 160 °C (FOX 200HT 160C)
-175 °C to 50 °C (FOX 200LT) |
Temperature Resolution |
±0.01 °C |
Accuracy |
±1% (Fox 200)
±2% (Fox 200HT 160C, FOX 200LT) |
Reproducibility |
±0.5% |
Thermal Conductivity Range |
0.005 to 0.35 W/m K
(0.035 to 2.4 BTU in/hr ft2 °F)
With FOX 200 External Thermocouple Kit:
0.001 to 2.5 W/m K
(0.007 to 17.3 BTU in/hr ft2 °F) |
Proprietary Thin Film
Heat Flux Transducer |
75 mm x 75 mm (3 in)
Top and bottom |